Letter of Credit at Sight (Sight L/C), Telegraphic Transfer (T/T)
100 Set Per Year
1 Months
Contact us for information regarding our sample policy
PLYWOOD
Asia, Central America, South America, Western Europe, Africa
All India
Product Description
Medical Face Mask Fire Resistance Tester is mainly used to test the flame retardant performance of medical masks. The combustion performance of the mask is measured through a head mold wearing a mask de-vice to contact with the flame at a certain linear speed. The device is mainly used in the field of medical protection, and also can also be used for testing flame retardant properties of other textile coating materials. Model:FTech-GB19083 Stabdard: GB 19083-2010 YY 0469-2011 GB 2626-2006 BS EN136-1998 FEATURES:
PLC + touch screen control system makes the test more convenient, operated and visualized;
The mask clamp is a metal human head mold, which can fully simulate the actual use state of the mask.
Burner height is adjustable.
Automatic timing and positioning of burner.
Combustion time and flame retardant time are recorded automatically and displayed digitally
Equipped with flame temperature measuring probe and flame temperature display.
Automatic control system, automatic ignition
The burner is equipped with LPG interface
Automatic timing, automatic storage of test data
Parameters
The linear velocity of automatic movement of metal head is (60 ± 5) mm / S.
Flame height is adjustable 40 ± 4mm
Equipped with thermocouple for flame temperature measurement and detection
The flame temperature is (800 ± 50) ºC at a height of (20 ± 2) mm from the top of the burner
The distance between the top of burner and the lowest point of mask is (20 ± 2) mm
Burner height adjustment range is 30mm
The thickness of metal head is 2-3mm which is fire resistant